Journal of Long-Term Effects of Medical Implants

DOI: 10.1615/JLongTermEffMedImplants.2014010563
pages 147-150

Severe Metallosis Following Polyethylene Dislocation in a Mobile-Bearing Medial Unicompartmental Knee Replacement

Alexandros P. Apostolopoulos
Trauma and Orthopaedic Department, "Korgialenio-Benakio" Hellenic Red Cross Hospital, Athens, Greece; Trauma and Orthopaedic Department, Ealing Hospital, North West University Healthcare NHS Trust, London, United Kingdom
Ilias Katsougrakis
Trauma and Orthopaedic Department, Ealing Hospital NHS Trust, London North West Healthcare, UK
Rafik Fanous
Trauma and Orthopaedic Department, Ealing Hospital, North West University Healthcare NHS Trust, London, United Kingdom
Andrew Harrison
Department of Trauma & Orthopaedics, Ealing Hospital NHS Trust, London, United Kingdom
Enrique Saavedra
Trauma and Orthopaedic Department, Ealing Hospital, North West University Healthcare NHS Trust, London, United Kingdom

SINOPSIS

Metallosis is a rare cause of failure after unicompartmental knee replacement (UKR) and has only previously been reported when there has been abnormal metal-on-metal contact. We present a case report of a 67-year-old woman who was admitted to our department with a 5-month history of a painful right knee. There was no history of trauma and her past medical history was unremarkable. She had undergone a primary mobile bearing medial UKR of her right knee 5 years ago. During the preceding 5 months, the patient complained that the knee was becoming unstable, was painful, and was making frequent clicking noises associated with metallic friction. Plain radiographs of the right knee showed dislocation of the polyethylene bearing and the patient was taken to the operating room for revision of the implant. Intraoperatively, severe evidence of metallosis was found in the periprosthetic soft tissues. The tibial component was found to be loose and the polyethylene bearing was dislocated posteriorly. The implants were revised. Postoperative recovery was uneventful. Microscopic examination revealed findings consistent with metallosis. At the latest follow-up, the patient had a good clinical and radiological outcome. In conclusion, such a complication of UKR requires early diagnosis and treatment as it can result in implant loosening and metallosis.


Articles with similar content:

Tibial Shaft Fracture after Removal of an ACE Tibial Nail
Journal of Long-Term Effects of Medical Implants, Vol.21, 2011, issue 3
loannis Sourlas, Stergios N. Lallos, Nikolaos E. Efstathopoulos, Michalis Papadakis, Emmanuel Brilakis
Trigeminal Neuralgia
Journal of Long-Term Effects of Medical Implants, Vol.16, 2006, issue 2
William B. Long III, Richard Edlich, Kathryne L. Winters, L. D. Britt
Retrieval Analysis of a Polycarbonate-Urethane Acetabular Cup: A Case Report
Journal of Long-Term Effects of Medical Implants, Vol.18, 2008, issue 1
Steven M. Kurtz
A Delayed Formation of Pretibial Pseudocyst After Anterior Cruciate Reconstruction from Bioabsorbable Interference Screws: A Report of Two Cases and Short Review of the Literature
Journal of Long-Term Effects of Medical Implants, Vol.28, 2018, issue 3
Marco Molinari, Avraam Christodoulidis, Mattia Barbareschi
Failure of a Medial Unicompartmental Knee Replacement due to Metal Allergy
Journal of Long-Term Effects of Medical Implants, Vol.28, 2018, issue 4
Shamsul Khan, Theodore Balfousias, George Tselentakis, Bala Srikanth, Alexandros P. Apostolopoulos