Suscripción a Biblioteca: Guest
Journal of Long-Term Effects of Medical Implants

Publicado 4 números por año

ISSN Imprimir: 1050-6934

ISSN En Línea: 1940-4379

SJR: 0.184 SNIP: 0.485 CiteScore™:: 1.1 H-Index: 39

Indexed in

Effect of Cement Type and Cementation Technique on the Retention of Implant-Supported Restorations

Volumen 30, Edición 1, 2020, pp. 61-67
DOI: 10.1615/JLongTermEffMedImplants.2020035290
Get accessGet access

SINOPSIS

Objectives: This study aimed to assess the effect of cement type and cementation technique on the retention of cement-retained implant-supported restorations.
Material and Methods: In this in vitro study, 20 solid abutments were screwed onto the fixture analogs and were mounted in acrylic resin molds using a surveyor. Twenty computer-aided design/computer-aided manufacturing (CAD/CAM) metal crowns with an occlusal loop were fabricated, divided into two groups (n = 10), and bonded with temporary dental cement (Temp-Bond; group 1) or glass ionomer (GC; group 2). Each group was subcategorized based on its cementation method (half-filling or abutment replica technique). The cement was applied to the copings that were placed vertically on the abutment with applying a 5-kg load for 10 min. The samples were then incubated at 37°C for 24 h and subjected to 3000 thermal cycles. The samples underwent axial tensile load within a universal testing machine. Data were analyzed using two-way ANOVA.
Results: GC cement yielded significantly higher retention than Temp-Bond (P < 0.05). In the GC group, the half-filling subgroup showed significantly higher retention than the abutment replica subgroup (P < 0.05). The mean retention value in the Temp-Bond group revealed no significant difference between the two subgroups (P > 0.05). GC yielded higher retention than Temp-Bond while using the half-filling cementation technique (P = 0.00). However, this difference was not significant between GC and Temp-Bond with the abutment replica technique (P = 0.960).
Conclusion: Within the limitations of this study, the results showed that permanent cementation with half-filling technique yielded significantly higher retention in implant restorations.

REFERENCIAS
  1. Quirynen M, Herrera D, Teughels W, Sanz M. Implant therapy: 40 years of experience. Periodontal 2000. 2014;66(1):7-12. .

  2. Canullo L, Cocchetto R, Marinotti F, Oltra DP, Diago MP, Loi I. Clinical evaluation of an improved cementation technique for implant-supported restorations: A randomized controlled trial. Clin Oral Implants Res. 2016;27(12):1492-9. .

  3. Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP. A systematic review of the 5-year survival and complication rates of implant-supported single crowns. Clin Oral Implants Res. 2008;19(2):119-30. .

  4. Gomez-Polo M, Ortega R, Gomez-Polo C, Celemin A, Del Rio Highsmith J. Factors affecting the decision to use cemented or screw-retained fixed implant-supported prostheses: A critical review. Int J Prosthodont. 2018;31(1):43-54. .

  5. Lee A, Okayasu K, Wang HL. Screw-versus cement-retained implant restorations: Current concepts. Implant Dent. 2010;19(1):8-15. .

  6. Wittneben JG, Millen C, Bragger U. Clinical performance of screw-versus cement-retained fixed implant-supported reconstructions-a systematic review. Int J Oral Maxillofac Implants. 2014;(29 Suppl):84-98. .

  7. Saber FS, Abolfazli N, Nuroloyuni S, Khodabakhsh S, Bahrami M, Nahidi R, Zeighami S. Effect of abutment height on retention of single cement-retained, wide- and narrow-platform implant-supported restorations. J Dent Res Dent Clin Dent Prospects. 2012;6(3):98-102. .

  8. Shadid R, Sadaqa N. A comparison between screw- and cement-retained implant prostheses. A literature review. J Oral Implantol. 2012;38(3):298-307. .

  9. Miyazaki T, Hotta Y, Kunii J, Kuriyama S, Tamaki Y. A review of dental CAD/CAM: Current status and future perspectives from 20 years of experience. Dent Mater J. 2009;28(1):44-56. .

  10. Mehl C, Harder S, Wolfart M, Kern M, Wolfart S. Retriev-ability of implant-retained crowns following cementation. Clin Oral Implants Res. 2008;19(12):1304-11. .

  11. Wittneben J-G, Millen C, Bragger U. Clinical performance of screw-versus cement-retained fixed implant-supported reconstructions-a systematic review. Int J Oral Maxillofac Implants. 2014;(29 Suppl):84-98. .

  12. Esposito M, Bressan E, Grusovin MG, D'Avenia F, Neumann K, Sbricoli L, Luongo G. Do repeated changes of abutments have any influence on the stability of peri-implant tissues? One-year post-loading results from a multi-centre randomised controlled trial. Eur J Oral Implantol. 2017;10(1):57-72. .

  13. Raees T. Cementation of implant prosthesis: A review of techniques for controlling excess cement around implants. Acta Sci Dent Sci. 2019;3(5):146-8. .

  14. Wilson TG Jr. The positive relationship between excess cement and peri-implant disease: A prospective clinical endoscopic study. J Periodontol. 2009;80(9):1388-92. .

  15. Nissan J, Snir D, Rosner O, Kolerman R, Chaushu L, Chaushu G. Reliability of retrievable cemented implant-supported prostheses. J Prosthet Dent. 2016;115(5): 587-91. .

  16. Millen C, Bragger U, Wittneben JG. Influence of prosthesis type and retention mechanism on complications with fixed implant-supported prostheses: A systematic review applying multivariate analyses. Int J Oral Maxillofac Implants. 2015;30(1):110-24. .

  17. Abduo J, Lyons K. Rationale for the use of CAD/ CAM technology in implant prosthodontics. Int J Dent. 2013;2013:768121. .

  18. Felton DA, Kanoy BE, White JT. The effect of surface roughness of crown preparations on retention of cemented castings. J Prosthet Dent. 1987;58(3):292-6. .

  19. GE K. Effect of surface treatments on implant crown re-tention. J Dent Sci. 2010;5:131-5. .

  20. Jorgensen KD. The relationship between retention and convergence angle in cemented veneer crowns. Acta Odontol Scand. 1955;13(1):35-40. .

  21. Kaufman EG, Coelho DH, Colin L. Factors influencing the retention of cemented gold castings. J Prosthet Dent. 1961;11:487-502. .

  22. Lewinstein I, Block L, Lehr Z, Ormianer Z, Matalon S. An in vitro assessment of circumferential grooves on the retention of cement-retained implant-supported crowns. J Prosthet Dent. 2011;106(6):367-72. .

  23. Miyazaki T, Hotta Y. CAD/CAM systems available for the fabrication of crown and bridge restorations. Aust Dent J. 2011;56(Suppl 1):97-106. .

  24. Carter GM, Hunter KM, Herbison P. Factors influencing the retention of cemented implant-supported crowns. N Z Dent J. 1997;93(412):36-8. .

  25. Abou-Obaid A, Al-Khudairy R. Effect of abutment height and cement thickness on the retention of cement-retained implant supported restorations. Adv Dent Oral Health. 2018; 9(4):00112-17. .

  26. Marin DOM, Leite ARP, Oliveira Junior NM, Paleari AG, Pero AC, Compagnoni MA. Retention force and wear characteristics of three attachment systems after dislodging cycles. Braz Dent J. 2018;29(6):576-82. .

  27. Saleh M, Tasar-Faruk S. Comparing the marginal leakage and retention of implant-supported restorations cemented by four different dental cements. Clin Implant Dent Relat Res. 2019;21(6):1181-8. .

  28. Amorfini L, Storelli S, Mosca D, Scanferla M, Romeo E. Comparison of cemented vs. screw-retained, customized computer-aided design/computer-assisted manufacture zirconia abutments for esthetically located single-tooth implants: A 10-year randomized prospective study. Int J Prosthodont. 2018;31(4):359-66. .

  29. Safari S, Hosseini Ghavam F, Amini P, Yaghmaei K. Effects of abutment diameter, luting agent type, and re-cementation on the retention of implant-supported CAD/ CAM metal copings over short abutments. J Adv Prosthodont. 2018;10(1):1-7. .

  30. Farzin M, Torabi K, Ahangari AH, Derafshi R. Effect of abutment modification and cement type on retention of cement-retained implant supported crowns. J Dent (Tehran). 2014;11(3):256-62. .

  31. Naumova EA, Roth F, Geis B, Baulig C, Arnold WH, Piwowarczyk A. Influence of luting materials on the retention of cemented implant-supported crowns: An in vitro study. Materials (Basel). 2018;11(10):1853. .

  32. Patel B, Patel V, Ramesh TR, Soundharya A. Effect of modified cementation technique on marginal fit and apical spread of excess cement for implant restorations: An in vitro study. J Dent Res Rev. 2016;3(1):8-12. .

  33. Alvarez-Arenal A, Gonzalez-Gonzalez I, de Llanos-Lanchares H, Brizuela-Velasco A, Pines-Hueso J, Ellakuria-Echebarria J. Retention strength after compressive cyclic loading of five luting agents used in implant-supported prostheses. Biomed Res Int. 2016;2016:2107027. .

  34. Saleh M, Faruk ST. Enhancing retrievability of cement retained implant supported restorations. EC Dent Sci. 2019;18(3):508-12. .

  35. Dudley JE, Richards LC, Abbott JR. Retention of cast crown copings cemented to implant abutments. Aust Dent J. 2008;53(4):332-9. .

1502 Vistas de artículos 42 Descargas de artículos Métrica
1502 PUNTOS DE VISTA 42 DESCARGAS Google
Scholar
CITAS

Artículos con contenido similar:

Aesthetic Outcome of Implants in the Anterior Zone after Socket Preservation and Conventional Implant Placement: A Retrospective Study Journal of Long-Term Effects of Medical Implants, Vol.30, 2020, issue 4
D. Rupawat, V. Sivaswamy, Subhabrata Maiti, D. Nallaswamy
Practice-Based Clinical Evaluation of Dental Implants Placed with Immediate or Delayed Loading Protocols: One-Year Follow-Up Journal of Long-Term Effects of Medical Implants, Vol.29, 2019, issue 2
Leonardo Rigoldi Bonjardim, Ernesto Byron Benalcazar Jalkh, Paulo Guilherme Coelho, Adolfo Coelho de Oliveira Lopes, Karin Hermana Neppelenbroek, Estevam Augusto Bonfante, Abbas Zahoui, Paulo Sérgio Perri de Carvalho de Carvalho
Effect of Bone Grafting on Hard and Soft Tissue Changes Following Immediate Implant Installation in Mandibular Molar Sites: A Double-Blind Randomized Clinical Trial Journal of Long-Term Effects of Medical Implants, Vol.33, 2023, issue 3
Kourosh Babaei, Nahid Nasrabadi, Mohammad Reza Farsi, Ali Forouzanfar, Farzaneh Ahrari
The Influence of Cortical Width of Sheep Bone on the Primary Stability of Dental Implants: An In Vitro Study Journal of Long-Term Effects of Medical Implants, Vol.23, 2013, issue 1
Amin Khodadoustan, Bijan Heidari, Mahdi Kadkhodazadeh, Meisam Khalesi
Comparison of Two Regenerative Surgical Treatments for Peri-Implantitis Defect using Natix Alone or in Combination with Bio-Oss and Collagen Membrane Journal of Long-Term Effects of Medical Implants, Vol.26, 2016, issue 3
Satlikh Mohamadi, Hamid Reza Arab, Amir Moeintaghavi, Najme Anbiaei, Farid Shiezadeh

Último edicion

Comparison of Malondialdehyde Levels among Patients with Sandblasted Acid-Etched and Anodized Surface Dental Implants: A Prospective Clinical Study Arvina Rajasekar, Sheeja S. Varghese Comparative Assessment of Catalase Levels in Peri-Implant Health and Disease M. Rithanya, Arvina Rajasekar Femoral Shaft Fracture Developed Late-Onset Infection in 27 Years after Intramedullary Nailing: A Case Report Yuichi Kasai, Ryo Okabayashi, Kentaro Aoki Comparative Evaluation of C-Reactive Protein Levels among Peri-Implant Health and Disease Conditions K. Sreekala Priyadharsini, Arvina Rajasekar Treatment of Schatzker Type III Tibial Plateau Fractures: Report of an Alternative, Percutaneous Technique and Brief Review of the Literature Avraam Christodoulidis, Piero Giardini, Carlo Raimondo Menna, Micaela Pagliari, Marco Molinari Ramus Frame Dental Implant Functional Survival and Safety in Severely Atrophic Edentulous Human Mandibles O. Hilt Tatum Jr., Thomas E. Rams The Role of Genicular Radiofrequency Ablation in the Management of Persistent Pain after Total Knee Arthroplasty Panagiotis Karampinas, Athanasios Galanis, Michail Vavourakis, Dimitrios Tzortzis, Evangelos Sakellariou, Dimitrios Zachariou, Spiros Karampitianis, John Vlamis Spiros Pneumaticos A Systematic Review on Maxillofacial Prosthesis with Respect to Their Color Stability Khushali K. Shah, Vaishnavi Rajaraman, Deepak Nallaswamy Veeraiyan, Subhabrata Maiti Robotic Arm-Assisted Total Knee Arthroplasty: Anatomical Alignment and Mid-Term Outcomes from the First Cohort Originating in Greece Konstantinos Dretakis, Vasilios G. Igoumenou, Symeon Naoum, Konstantinos Raptis, Christos Koutserimpas Relationship between Impacted Mandibular Third Molars and the Mandibular Canal on CBCT Scans Yaser Safi, Mahkameh Moshfeghi, Mitra Ghazizadeh Ahsaie, Maryam Zameni, Seyed Ahmadreza Sahafi Association of Age, Gender, and Site of Implant Placement: An Institution-Based Retrospective Study Harini K. Sri, Nabeel Ahmed, L. Keerthi Sasanka The Impact of Patient Characteristics on Outcomes of Surgically Managed Vertebral Osteomyelitis in the United States: Insights from a National Database Study Jennifer Kurowicki, Stuart Changoor, Daniel Coban, Neil Patel, Kumar Sinha, Ki Hwang, Arash Emami

Próximos Artículos

Transfer of guided emergence profile developed using customized provisional implant restoration and cementation using an abutment replica technique- A case report. Vinay Sivaswamy, Balaji Anand S D, Jacob Mathew Philip, C J Venkatakrishnan, Narasimman M, Helen Mary Abraham, Pavithra B Marijuana use is Associated with Increased Rates of Hip Dislocation and Lower Insurance Reimbursement among THA Recipients Andrew Recker, Chukwuweike Gwam, Erin Kelly, Scott Douglas, Johannes Plate Surface modification of Mg alloy AZ91D using nanopowder mixed EDM for Biodegradable Implant Alok Kumar, Abhishek Singh The history of orthopedics in the long term: from general medicine to medical specialties Pavlos Altsitzioglou, Ioannis Zafeiris, Fotini Soucacos, Pantelis Limneos, Eleftheria Souliotis, Evanthia Mitsiokapa, Theodosis Saranteas, Andreas Mavrogenis COMPARISON OF ROUGHNESS, WETTABILITY AND SCANNING ELECTRON MICROSCOPY FEATURES BETWEEN SANDBLASTED ACID-ETCHED AND OXIDIZED TITANIUM DENTAL IMPLANTS Kshitiz Chhabra, Arvina Rajasekar CORRELATION BETWEEN PRIMARY STABILITY OF DENTAL IMPLANTS AND BONE DENSITY: A RETROSPECTIVE ANALYSIS Padmaja S, Arvina Rajasekar C-Reactive Protein and Erythrocyte Sedimentation Rates after total and unicompartmental knee arthroplasty - less implant equals quicker normalization Kasım Kılıçarslan, Ömer Faruk Naldöven, Enejd Veizi, Şahan Güven, Şahin Çepni, Ahmet Fırat COMPLICATIONS FOLLOWING SHORT FEMORAL NAIL FIXATION FOR INTETROCHANTERIC HIP FRACTURES. A RETROSPECTIVE STUDY Anastasia Vasilopoulou, Antonios Trichonas, Konstantinos Palaiologos, Emmanouel Antonogiannakis, Christos Nikakis, Stavros Angelis, Spyridon Maris, Christos Grekas , Efthimios Karadimas, Alexandros Apostolopoulos FEA: Connector Designs and Pontic Stress Distribution of Fixed Partial Denture Implant-Supported Metal Framework Saja Muhsin, Enas Mohammed, Khalid Bander Design and Frequency Stability Analysis of an Adaptive Neuro-Fuzzy Inference System based Artificial Pacemaker Controller in MATLAB Nader Jafarnia Dabanloo, Asghar Dabiriaghdam, Fereidoon Nooshirvan Rahatabad, Keivan Maghooli Complications following intraosseous injections of calcium phosphate bone cement in Subchondroplasty. Christos Zachariadis, Helen Leligou, Stauros Kourkoulis, Evaggelos Magnisalis, Panayiotis Papagelopoulos, Olga Savvidou The Impact of Patient Characteristics on Outcomes of Surgically Managed Vertebral Osteomyelitis in the United States: Insights from a National Database Study Jennifer Kurowicki, Stuart Changoor, Daniel Coban, Neil Patel, Kumar Sinha, Ki Hwang, Arash Emami ASSOCIATION OF AGE, GENDER AND SITE OF IMPLANT PLACEMENT - AN INSTITUTION BASED RETROSPECTIVE STUDY Harini Sri, Nabeel Ahmed, Keerthi Shasanka
Portal Digitalde Biblioteca Digital eLibros Revistas Referencias y Libros de Ponencias Colecciones Precios y Políticas de Suscripcione Begell House Contáctenos Language English 中文 Русский Português German French Spain