Library Subscription: Guest
Journal of Long-Term Effects of Medical Implants

Published 4 issues per year

ISSN Print: 1050-6934

ISSN Online: 1940-4379

SJR: 0.184 SNIP: 0.485 CiteScore™:: 1.1 H-Index: 39

Indexed in

Early Fracture of the Trunion in Total Hip Arthroplasty: A Case Report

Volume 29, Issue 3, 2019, pp. 183-186
DOI: 10.1615/JLongTermEffMedImplants.2019032624
Get accessGet access

ABSTRACT

A 59-year-old man who had previously undergone total hip arthroplasty (THA) with the use of a dual modular (head and neck) total hip implant presented with a mechanical failure at the trunion 9 yr after index surgery with a low-energy mechanism. The fractured stem was then removed and a revision stem implanted to restore the patient's ability to ambulate. We demonstrate and contribute to the small but growing evidence of failure of modular THA systems at the trunion.

REFERENCES
  1. Wright G, Sporer S, Urban R, Jacobs J. Fracture of a modular femoral neck after total hip arthroplasty: a case report. J Bone Joint Surg Am. 2010;92(6):1518-21. .

  2. Ellman MB, Levine BR. Fracture of the modular femoral neck component in total hip arthroplasty. J Arthroplasty. 2013;28(1):196.e1-5. .

  3. Wodecki P, Sabbah D, Kermarrec G, Semaan I. New type of hip arthroplasty failure related to modular femoral components: breakage at the neck-stem junction. Orthop Traumatol Surg Res. 2013;99(6):741-4. .

  4. Morley D, Starks I, Lim J. A case of a C-stem fracture at the head-neck junction and a review of the literature. Case Rep Orthop. 2012;158604. doi: 10.1155/2012/158604. .

  5. Skendzel JG, Blaha JD, Urquhart AG. Total hip arthro-plasty modular neck failure. J Arthroplasty. 2011;26(2): 338.e1-4. .

  6. Botti TP, Gent J, Martell JM, Manning DW. Trunion frac-ture of a fully porous-coated femoral stem. Case report. J Arthroplasty. 2005;20(7):943-5. .

  7. Wilson DA, Dunbar MJ, Amirault JD, Farhat Z. Early failure of a modular femoral neck total hip arthroplasty component: a case report. J Bone Joint Surg Am. 2010;92(6):1514-7. .

  8. Brown SA, Flemming CA, Kawalec JS, Placko HE, Vassaux C, Merritt K, Payer JH, Kraay MJ. Fretting corrosion accelerates crevice corrosion of modular hip tapers. J Appl Biomater. 1995;6(1):19-26. .

  9. Kwon YM, Antoci V, Eisemon E, Tsai TY, Yan Y, Liow MH. "Top-out" removal of well-fixed dual-taper femoral stems: surgical technique and radiographic risk factors. J Arthroplasty. 2016;31(12):2843-9. .

  10. Pivec R, Meneghini RM, Hozack WJ, Westrich GH, Mont MA. Modular taper junction corrosion and failure: how to approach a recalled total hip arthroplasty implant. J Arthroplasty. 2014;29(1):1-6..

762 Article views 12 Article downloads Metrics
762 VIEWS 12 DOWNLOADS Google
Scholar
CITATIONS

Articles with similar content:

Early Fracture of the Modular Femoral Neck Component in Total Hip Arthroplasty: A Case Report Journal of Long-Term Effects of Medical Implants, Vol.28, 2018, issue 4
Che Hang Jason Wong, Angelo Mannino
Failure of a Medial Unicompartmental Knee Replacement due to Metal Allergy Journal of Long-Term Effects of Medical Implants, Vol.28, 2018, issue 4
Shamsul Khan, Theodore Balfousias, George Tselentakis, Bala Srikanth, Alexandros P. Apostolopoulos
Complications Following Short Femoral Nail Fixation for Intertrochanteric Hip Fractures: A Retrospective Study Journal of Long-Term Effects of Medical Implants, Vol.34, 2024, issue 4
Konstantinos Palaiologos, Spyridon J. Maris, Anastasia Vasilopoulou, Stavros Angelis, Antonios Trichonas, C. Grekas, Emmanouel Antonogiannakis, Alexandros P. Apostolopoulos, C. Nikakis, Efthymios J. Karadimas
High Survivorship of Hybrid Fixation Technique in Aseptic Condylar Revision Total Knee Arthroplasty with Minimal Metaphyseal Bone Loss: 5−10 Year Clinical Outcomes Journal of Long-Term Effects of Medical Implants, Vol.34, 2024, issue 1
Hosam E. Matar, Jeffrey D. Gollish, Rajesh Bawale
Case Report: Pseudotumor associated with corrosion of a femoral component with a modular neck and a ceramic-on-polyethylene bearing Journal of Long-Term Effects of Medical Implants, Vol.24, 2014, issue 1
Stuart B. Goodman, Joe Messana, Muyibat Adelani

Forthcoming Articles

Transfer of guided emergence profile developed using customized provisional implant restoration and cementation using an abutment replica technique- A case report. Vinay Sivaswamy, Balaji Anand S D, Jacob Mathew Philip, C J Venkatakrishnan, Narasimman M, Helen Mary Abraham, Pavithra B Marijuana use is Associated with Increased Rates of Hip Dislocation and Lower Insurance Reimbursement among THA Recipients Andrew Recker, Chukwuweike Gwam, Erin Kelly, Scott Douglas, Johannes Plate Surface modification of Mg alloy AZ91D using nanopowder mixed EDM for Biodegradable Implant Alok Kumar, Abhishek Singh The history of orthopedics in the long term: from general medicine to medical specialties Pavlos Altsitzioglou, Ioannis Zafeiris, Fotini Soucacos, Pantelis Limneos, Eleftheria Souliotis, Evanthia Mitsiokapa, Theodosis Saranteas, Andreas Mavrogenis COMPARISON OF ROUGHNESS, WETTABILITY AND SCANNING ELECTRON MICROSCOPY FEATURES BETWEEN SANDBLASTED ACID-ETCHED AND OXIDIZED TITANIUM DENTAL IMPLANTS Kshitiz Chhabra, Arvina Rajasekar CORRELATION BETWEEN PRIMARY STABILITY OF DENTAL IMPLANTS AND BONE DENSITY: A RETROSPECTIVE ANALYSIS Padmaja S, Arvina Rajasekar C-Reactive Protein and Erythrocyte Sedimentation Rates after total and unicompartmental knee arthroplasty - less implant equals quicker normalization Kasım Kılıçarslan, Ömer Faruk Naldöven, Enejd Veizi, Şahan Güven, Şahin Çepni, Ahmet Fırat COMPLICATIONS FOLLOWING SHORT FEMORAL NAIL FIXATION FOR INTETROCHANTERIC HIP FRACTURES. A RETROSPECTIVE STUDY Anastasia Vasilopoulou, Antonios Trichonas, Konstantinos Palaiologos, Emmanouel Antonogiannakis, Christos Nikakis, Stavros Angelis, Spyridon Maris, Christos Grekas , Efthimios Karadimas, Alexandros Apostolopoulos FEA: Connector Designs and Pontic Stress Distribution of Fixed Partial Denture Implant-Supported Metal Framework Saja Muhsin, Enas Mohammed, Khalid Bander Design and Frequency Stability Analysis of an Adaptive Neuro-Fuzzy Inference System based Artificial Pacemaker Controller in MATLAB Nader Jafarnia Dabanloo, Asghar Dabiriaghdam, Fereidoon Nooshirvan Rahatabad, Keivan Maghooli Complications following intraosseous injections of calcium phosphate bone cement in Subchondroplasty. Christos Zachariadis, Helen Leligou, Stauros Kourkoulis, Evaggelos Magnisalis, Panayiotis Papagelopoulos, Olga Savvidou The Impact of Patient Characteristics on Outcomes of Surgically Managed Vertebral Osteomyelitis in the United States: Insights from a National Database Study Jennifer Kurowicki, Stuart Changoor, Daniel Coban, Neil Patel, Kumar Sinha, Ki Hwang, Arash Emami ASSOCIATION OF AGE, GENDER AND SITE OF IMPLANT PLACEMENT - AN INSTITUTION BASED RETROSPECTIVE STUDY Harini Sri, Nabeel Ahmed, Keerthi Shasanka