Effect of Cement Type and Cementation Technique on the Retention of Implant-Supported Restorations

巻 30, 発行 1, 2020, pp. 61-67
DOI: 10.1615/JLongTermEffMedImplants.2020035290
Get accessGet access

要約

Objectives: This study aimed to assess the effect of cement type and cementation technique on the retention of cement-retained implant-supported restorations.
Material and Methods: In this in vitro study, 20 solid abutments were screwed onto the fixture analogs and were mounted in acrylic resin molds using a surveyor. Twenty computer-aided design/computer-aided manufacturing (CAD/CAM) metal crowns with an occlusal loop were fabricated, divided into two groups (n = 10), and bonded with temporary dental cement (Temp-Bond; group 1) or glass ionomer (GC; group 2). Each group was subcategorized based on its cementation method (half-filling or abutment replica technique). The cement was applied to the copings that were placed vertically on the abutment with applying a 5-kg load for 10 min. The samples were then incubated at 37°C for 24 h and subjected to 3000 thermal cycles. The samples underwent axial tensile load within a universal testing machine. Data were analyzed using two-way ANOVA.
Results: GC cement yielded significantly higher retention than Temp-Bond (P < 0.05). In the GC group, the half-filling subgroup showed significantly higher retention than the abutment replica subgroup (P < 0.05). The mean retention value in the Temp-Bond group revealed no significant difference between the two subgroups (P > 0.05). GC yielded higher retention than Temp-Bond while using the half-filling cementation technique (P = 0.00). However, this difference was not significant between GC and Temp-Bond with the abutment replica technique (P = 0.960).
Conclusion: Within the limitations of this study, the results showed that permanent cementation with half-filling technique yielded significantly higher retention in implant restorations.

参考
  1. Quirynen M, Herrera D, Teughels W, Sanz M. Implant therapy: 40 years of experience. Periodontal 2000. 2014;66(1):7-12. .

  2. Canullo L, Cocchetto R, Marinotti F, Oltra DP, Diago MP, Loi I. Clinical evaluation of an improved cementation technique for implant-supported restorations: A randomized controlled trial. Clin Oral Implants Res. 2016;27(12):1492-9. .

  3. Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP. A systematic review of the 5-year survival and complication rates of implant-supported single crowns. Clin Oral Implants Res. 2008;19(2):119-30. .

  4. Gomez-Polo M, Ortega R, Gomez-Polo C, Celemin A, Del Rio Highsmith J. Factors affecting the decision to use cemented or screw-retained fixed implant-supported prostheses: A critical review. Int J Prosthodont. 2018;31(1):43-54. .

  5. Lee A, Okayasu K, Wang HL. Screw-versus cement-retained implant restorations: Current concepts. Implant Dent. 2010;19(1):8-15. .

  6. Wittneben JG, Millen C, Bragger U. Clinical performance of screw-versus cement-retained fixed implant-supported reconstructions-a systematic review. Int J Oral Maxillofac Implants. 2014;(29 Suppl):84-98. .

  7. Saber FS, Abolfazli N, Nuroloyuni S, Khodabakhsh S, Bahrami M, Nahidi R, Zeighami S. Effect of abutment height on retention of single cement-retained, wide- and narrow-platform implant-supported restorations. J Dent Res Dent Clin Dent Prospects. 2012;6(3):98-102. .

  8. Shadid R, Sadaqa N. A comparison between screw- and cement-retained implant prostheses. A literature review. J Oral Implantol. 2012;38(3):298-307. .

  9. Miyazaki T, Hotta Y, Kunii J, Kuriyama S, Tamaki Y. A review of dental CAD/CAM: Current status and future perspectives from 20 years of experience. Dent Mater J. 2009;28(1):44-56. .

  10. Mehl C, Harder S, Wolfart M, Kern M, Wolfart S. Retriev-ability of implant-retained crowns following cementation. Clin Oral Implants Res. 2008;19(12):1304-11. .

  11. Wittneben J-G, Millen C, Bragger U. Clinical performance of screw-versus cement-retained fixed implant-supported reconstructions-a systematic review. Int J Oral Maxillofac Implants. 2014;(29 Suppl):84-98. .

  12. Esposito M, Bressan E, Grusovin MG, D'Avenia F, Neumann K, Sbricoli L, Luongo G. Do repeated changes of abutments have any influence on the stability of peri-implant tissues? One-year post-loading results from a multi-centre randomised controlled trial. Eur J Oral Implantol. 2017;10(1):57-72. .

  13. Raees T. Cementation of implant prosthesis: A review of techniques for controlling excess cement around implants. Acta Sci Dent Sci. 2019;3(5):146-8. .

  14. Wilson TG Jr. The positive relationship between excess cement and peri-implant disease: A prospective clinical endoscopic study. J Periodontol. 2009;80(9):1388-92. .

  15. Nissan J, Snir D, Rosner O, Kolerman R, Chaushu L, Chaushu G. Reliability of retrievable cemented implant-supported prostheses. J Prosthet Dent. 2016;115(5): 587-91. .

  16. Millen C, Bragger U, Wittneben JG. Influence of prosthesis type and retention mechanism on complications with fixed implant-supported prostheses: A systematic review applying multivariate analyses. Int J Oral Maxillofac Implants. 2015;30(1):110-24. .

  17. Abduo J, Lyons K. Rationale for the use of CAD/ CAM technology in implant prosthodontics. Int J Dent. 2013;2013:768121. .

  18. Felton DA, Kanoy BE, White JT. The effect of surface roughness of crown preparations on retention of cemented castings. J Prosthet Dent. 1987;58(3):292-6. .

  19. GE K. Effect of surface treatments on implant crown re-tention. J Dent Sci. 2010;5:131-5. .

  20. Jorgensen KD. The relationship between retention and convergence angle in cemented veneer crowns. Acta Odontol Scand. 1955;13(1):35-40. .

  21. Kaufman EG, Coelho DH, Colin L. Factors influencing the retention of cemented gold castings. J Prosthet Dent. 1961;11:487-502. .

  22. Lewinstein I, Block L, Lehr Z, Ormianer Z, Matalon S. An in vitro assessment of circumferential grooves on the retention of cement-retained implant-supported crowns. J Prosthet Dent. 2011;106(6):367-72. .

  23. Miyazaki T, Hotta Y. CAD/CAM systems available for the fabrication of crown and bridge restorations. Aust Dent J. 2011;56(Suppl 1):97-106. .

  24. Carter GM, Hunter KM, Herbison P. Factors influencing the retention of cemented implant-supported crowns. N Z Dent J. 1997;93(412):36-8. .

  25. Abou-Obaid A, Al-Khudairy R. Effect of abutment height and cement thickness on the retention of cement-retained implant supported restorations. Adv Dent Oral Health. 2018; 9(4):00112-17. .

  26. Marin DOM, Leite ARP, Oliveira Junior NM, Paleari AG, Pero AC, Compagnoni MA. Retention force and wear characteristics of three attachment systems after dislodging cycles. Braz Dent J. 2018;29(6):576-82. .

  27. Saleh M, Tasar-Faruk S. Comparing the marginal leakage and retention of implant-supported restorations cemented by four different dental cements. Clin Implant Dent Relat Res. 2019;21(6):1181-8. .

  28. Amorfini L, Storelli S, Mosca D, Scanferla M, Romeo E. Comparison of cemented vs. screw-retained, customized computer-aided design/computer-assisted manufacture zirconia abutments for esthetically located single-tooth implants: A 10-year randomized prospective study. Int J Prosthodont. 2018;31(4):359-66. .

  29. Safari S, Hosseini Ghavam F, Amini P, Yaghmaei K. Effects of abutment diameter, luting agent type, and re-cementation on the retention of implant-supported CAD/ CAM metal copings over short abutments. J Adv Prosthodont. 2018;10(1):1-7. .

  30. Farzin M, Torabi K, Ahangari AH, Derafshi R. Effect of abutment modification and cement type on retention of cement-retained implant supported crowns. J Dent (Tehran). 2014;11(3):256-62. .

  31. Naumova EA, Roth F, Geis B, Baulig C, Arnold WH, Piwowarczyk A. Influence of luting materials on the retention of cemented implant-supported crowns: An in vitro study. Materials (Basel). 2018;11(10):1853. .

  32. Patel B, Patel V, Ramesh TR, Soundharya A. Effect of modified cementation technique on marginal fit and apical spread of excess cement for implant restorations: An in vitro study. J Dent Res Rev. 2016;3(1):8-12. .

  33. Alvarez-Arenal A, Gonzalez-Gonzalez I, de Llanos-Lanchares H, Brizuela-Velasco A, Pines-Hueso J, Ellakuria-Echebarria J. Retention strength after compressive cyclic loading of five luting agents used in implant-supported prostheses. Biomed Res Int. 2016;2016:2107027. .

  34. Saleh M, Faruk ST. Enhancing retrievability of cement retained implant supported restorations. EC Dent Sci. 2019;18(3):508-12. .

  35. Dudley JE, Richards LC, Abbott JR. Retention of cast crown copings cemented to implant abutments. Aust Dent J. 2008;53(4):332-9. .

1517 記事の閲覧数 42 記事のダウンロード 記事の統計
1517 記事の閲覧数 42 記事のダウンロード Google
Scholar
引用数

類似内容の記事:

Practice-Based Clinical Evaluation of Dental Implants Placed with Immediate or Delayed Loading Protocols: One-Year Follow-Up Journal of Long-Term Effects of Medical Implants, Vol.29, 2019, issue 2
Leonardo Rigoldi Bonjardim, Ernesto Byron Benalcazar Jalkh, Paulo Guilherme Coelho, Adolfo Coelho de Oliveira Lopes, Karin Hermana Neppelenbroek, Estevam Augusto Bonfante, Abbas Zahoui, Paulo Sérgio Perri de Carvalho de Carvalho
Radiographic Evaluation of Bone-Level Implants with Wound Dehiscence between the First- and Second-Stage Surgeries Journal of Long-Term Effects of Medical Implants, Vol.26, 2016, issue 3
Mohsen Bidgoli, Javad Faradmal, Sara Soheilifar, Erfan Abbasi Atibeh
Evaluation of Marginal and Internal Fit at Implant-Abutment Interface of Original and Compatible Nonoriginal Abutments Journal of Long-Term Effects of Medical Implants, Vol.32, 2022, issue 3
Bouchra Hjeij, Georges Aoun, Antoine Berberi, Georges Tehini
Whole-body Vibration Training in Older Adults: Retention of the Strengthening Effects Critical Reviews™ in Physical and Rehabilitation Medicine, Vol.24, 2012, issue 1-2
Sean Horton, Nancy McNevin, Chantelle C. Lachance, Kelly M. Carr, Patricia L. Weir, Kenji A. Kenno
Comparison of the Acceptability of a New Scoring System with Misch's Classification for Dental Implant Success Determination Journal of Long-Term Effects of Medical Implants, Vol.22, 2012, issue 1
Parisa Zarnegarnia, Zeinab Rezaei Esfahrood, Mahdi Kadkhodazadeh, Reza Amid

近刊の記事

Transfer of guided emergence profile developed using customized provisional implant restoration and cementation using an abutment replica technique- A case report. Vinay Sivaswamy, Balaji Anand S D, Jacob Mathew Philip, C J Venkatakrishnan, Narasimman M, Helen Mary Abraham, Pavithra B Marijuana use is Associated with Increased Rates of Hip Dislocation and Lower Insurance Reimbursement among THA Recipients Andrew Recker, Chukwuweike Gwam, Erin Kelly, Scott Douglas, Johannes Plate Surface modification of Mg alloy AZ91D using nanopowder mixed EDM for Biodegradable Implant Alok Kumar, Abhishek Singh The history of orthopedics in the long term: from general medicine to medical specialties Pavlos Altsitzioglou, Ioannis Zafeiris, Fotini Soucacos, Pantelis Limneos, Eleftheria Souliotis, Evanthia Mitsiokapa, Theodosis Saranteas, Andreas Mavrogenis COMPARISON OF ROUGHNESS, WETTABILITY AND SCANNING ELECTRON MICROSCOPY FEATURES BETWEEN SANDBLASTED ACID-ETCHED AND OXIDIZED TITANIUM DENTAL IMPLANTS Kshitiz Chhabra, Arvina Rajasekar CORRELATION BETWEEN PRIMARY STABILITY OF DENTAL IMPLANTS AND BONE DENSITY: A RETROSPECTIVE ANALYSIS Padmaja S, Arvina Rajasekar C-Reactive Protein and Erythrocyte Sedimentation Rates after total and unicompartmental knee arthroplasty - less implant equals quicker normalization Kasım Kılıçarslan, Ömer Faruk Naldöven, Enejd Veizi, Şahan Güven, Şahin Çepni, Ahmet Fırat COMPLICATIONS FOLLOWING SHORT FEMORAL NAIL FIXATION FOR INTETROCHANTERIC HIP FRACTURES. A RETROSPECTIVE STUDY Anastasia Vasilopoulou, Antonios Trichonas, Konstantinos Palaiologos, Emmanouel Antonogiannakis, Christos Nikakis, Stavros Angelis, Spyridon Maris, Christos Grekas , Efthimios Karadimas, Alexandros Apostolopoulos FEA: Connector Designs and Pontic Stress Distribution of Fixed Partial Denture Implant-Supported Metal Framework Saja Muhsin, Enas Mohammed, Khalid Bander Design and Frequency Stability Analysis of an Adaptive Neuro-Fuzzy Inference System based Artificial Pacemaker Controller in MATLAB Nader Jafarnia Dabanloo, Asghar Dabiriaghdam, Fereidoon Nooshirvan Rahatabad, Keivan Maghooli Complications following intraosseous injections of calcium phosphate bone cement in Subchondroplasty. Christos Zachariadis, Helen Leligou, Stauros Kourkoulis, Evaggelos Magnisalis, Panayiotis Papagelopoulos, Olga Savvidou The Impact of Patient Characteristics on Outcomes of Surgically Managed Vertebral Osteomyelitis in the United States: Insights from a National Database Study Jennifer Kurowicki, Stuart Changoor, Daniel Coban, Neil Patel, Kumar Sinha, Ki Hwang, Arash Emami ASSOCIATION OF AGE, GENDER AND SITE OF IMPLANT PLACEMENT - AN INSTITUTION BASED RETROSPECTIVE STUDY Harini Sri, Nabeel Ahmed, Keerthi Shasanka
Begell Digital Portal Begellデジタルライブラリー 電子書籍 ジャーナル 参考文献と会報 リサーチ集 価格及び購読のポリシー Begell House 連絡先 Language English 中文 Русский Português German French Spain