pages 13-19

Fourteen-Year Prospective Results of a High-Density Polyethylene Prosthetic Anterior Cruciate Ligament Reconstruction

Robert Purchase
Temple University Hospital, Philadelphia, PA
Richard Mason
The Orthopedic Center, Easton, MD
Victor Hsu
Temple University Hospital, Philadelphia, PA
Kenneth Rogers
Temple University Hospital, Philadelphia, PA
John P. Gaughan
Joseph Torg
Temple University Hospital, Philadelphia, PA

RESUMO

Our experience with an ultrahigh molecular weight polyethylene (UHMWPE) braided graft indicates that this prosthetic anterior cruciate ligament (ACL) is not associated with the complications of other prosthetic ligaments and has equivalent results compared to an autograft. Nine patients underwent ACL reconstruction with the UHMWPE prosthetic graft, and seven patients received a bone-patellar-bone autograft. Each patient completed a questionnaire regarding their symptoms, a Tegner score, and a Lysholm Score. Objective results included a physical examination and KT1000 measurement. Noninferiority statistical methods were utilized to calculate a power to detect a clinically relevant difference between the groups. Six individuals from the prosthetic group and six of the controls have been followed an average of 14.1 years. In the prosthetic group, one graft failed at 9.5 years after the reconstruction. There were no sterile effusions. There were no graft failures or sterile effusions in the controls. Tegner scores were equivalent (control = 5.5, prosthetic = 5.0) as were the Lysholm scores (control = 82, prosthetic = 80.5). KT1000 testing for the prosthetic group (mean side-to-side difference in maximal displacement = 1.75mm) was not inferior to the control group (2.0mm). In this prospective study with 14 years of follow-up, the subjective and objective results of a prosthetic ACL compares favorably to an autograft, suggesting reconsideration of a prosthetic graft options as a viable option.


Articles with similar content:

Analysis of Spinal Kinematics Following Implantation of Lumbar Spine Disc Prostheses versus Fusion: Radiological Study
Journal of Long-Term Effects of Medical Implants, Vol.17, 2007, issue 3
B. Ishak, O. Kollmar, O. Steimer, W. I. Steudel, A. Al-Yhary, Abdullah Nabhan
Effectiveness of Negative-Pressure Wound Therapy following Total Hip and Knee Replacements
Journal of Long-Term Effects of Medical Implants, Vol.29, 2019, issue 1
Videshnandan Raut, Hosam E. Matar, Nicholas Emms
Sterile Pretibial Cyst Formation Following Anterior Cruciate Ligament Reconstruction With a Bioabsorbable Screw
Journal of Long-Term Effects of Medical Implants, Vol.23, 2013, issue 1
Jonathan Riguad, Eric Tan, Johnathan A. Bernard, Bashir A. Zikria, Qais Naziri
Hip Pain and Pseudo-Lengthening of the Leg Due to Iliopsoas Haematoma Following Implantation of an Uncemented Component in Acetabular Cotyloplasty Technique
Journal of Long-Term Effects of Medical Implants, Vol.22, 2012, issue 4
Raghavendra Prasad Sidaginamale, A. Manta, B. McConnell, Alexandros P. Apostolopoulos, G. Zafiropoulos
The Outcome and Complications of ACL Reconstruction Augmented with a Ligament Augmentation Device for a Follow-up Period of More Than Five years
Journal of Long-Term Effects of Medical Implants, Vol.10, 2000, issue 3
Ichiro Sekiya, Yusuke Sakaguchi, Haruyasu Yamamoto, Kenichi Shinomiya, Takeshi Muneta