Neutral Temperature Range in Incubators: Performance of Equipment in Current Use and New Developments

Том 25, Выпуск 4-5, 1997, pp. 287-370
DOI: 10.1615/CritRevBiomedEng.v25.i4-5.10
Get accessGet access

Краткое описание

Low-birth-weight neonates should be nursed at thermoneutrality inside incubators. Thermoneutrality control is essential to enhance body growth and to reduce neonatal illnesses and mortality. Guidelines have been published to provide the thermoneutral range, but the recommendations did not always take into account all ambient and physiological parameters influencing thermoneutrality. In most marketed incubators, the heat supply is controlled through convective air flow (closed incubators) or through radiant power density (radiant warmer beds). The heating unit (on/off cycling or adjustable proportional control) is activated by an error signal calculated from the difference between a controlled temperature and a reference value preset by the clinician. The controlled variable can be either the incubator air temperature or the skin temperature of the anterior abdominal region of the neonate.
The neonate's size, thermal properties of the mattress and of incubator walls, air temperature and humidity, air velocity, incubator wall temperatures all influence the heat exchanges between the neonate and the surroundings, and, consequently, modify the obtention of thermoneutrality. Moreover, studies of the physiological mechanisms by which the neonate regulates body heat storage suggest that metabolic rate, behavior, vigilance level, nursing care, and heater control processes should also be taken into account. Little attention has been paid to these factors, and incubator performances are often disappointing.
This article reviews the different factors that modify thermoneutral condition. An attempt is made to suggest new ways to design equipment incorporating these factors in algorithms controlling heater processes in order to reach the optimal thermal environment in which the neonate should be nursed.

ЦИТИРОВАНО В
  1. Kim Y. -H., Kwon C. -H., Yoo S. -C., Experimental and numerical studies on convective heat transfer in a neonatal incubator, Medical & Biological Engineering & Computing, 40, 1, 2002. Crossref

  2. Bissinger Robin L., Annibale David J., Thermoregulation in Very Low-Birth-Weight Infants During the Golden Hour, Advances in Neonatal Care, 10, 5, 2010. Crossref

  3. Deguines C, Dégrugilliers L, Ghyselen L, Chardon K, Bach V, Tourneux P, Impact of nursing care on temperature environment in preterm newborns nursed in closed convective incubators, Acta Paediatrica, 102, 3, 2013. Crossref

  4. Abbas Abbas K., Heiman Konrad, Jergus Katrin, Orlikowsky Thorsten, Leonhardt Steffen, Neonatal Infrared Thermography Monitoring, in Neonatal Monitoring Technologies, 2012. Crossref

  5. Khodadadeh Yassaman, Comparative Evaluation of 2 Containers for a Reusable Heat Source Used in a Nonelectric Neonatal Transport Incubator, Journal of Clinical Engineering, 32, 2, 2007. Crossref

  6. Carpentier E., Décima P., Dégrugilliers L., Delanaud S., Bach V., Libert J. -P., Tourneux P., Les températures cutanées locales mesurées en pratique clinique sont-elles prédictives de la température cutanée moyenne chez le nouveau-né prématuré ?, Revue de médecine périnatale, 5, 4, 2013. Crossref

  7. Belghazi Khalid, Tourneux Pierre, Elabbassi Elmountacer Billah, Ghyselen Laurent, Delanaud Stephane, Libert Jean-Pierre, Effect of posture on the thermal efficiency of a plastic bag wrapping in neonate: Assessment using a thermal “sweating” mannequin, Medical Physics, 33, 3, 2006. Crossref

  8. Ghyselen Laurent, Fontaine Cécile, Dégrugilliers Loïc, Degorre Coralie, Léké Andre, Tourneux Pierre, Polyethylene bag wrapping to prevent hypothermia during percutaneous central venous catheter insertion in the preterm newborn under 32 weeks of gestation, The Journal of Maternal-Fetal & Neonatal Medicine, 27, 18, 2014. Crossref

  9. Uystepruyst C.H., Coghe J., Dorts T.H., Harmegnies N., Delsemme M.H., Art T., Lekeux P., Effect of Three Resuscitation Procedures on Respiratory and Metabolic Adaptation to Extra Uterine Life in Newborn Calves, The Veterinary Journal, 163, 1, 2002. Crossref

  10. Amorim M.F., Oliveira G.H.C., Silva D.S.R., Kiapuchinski L.M., Water vapour partial pressure control in neonatal incubator using prevision by multiple models, The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 3, 2004. Crossref

  11. Deguines C, Décima P, Pelletier A, Dégrugilliers L, Ghyselen L, Tourneux P, Variations in incubator temperature and humidity management: a survey of current practice, Acta Paediatrica, 101, 3, 2012. Crossref

  12. Telliez F., Bach V., Delanaud S., M'Baye H., Leke A., Apédoh A., Abdiche M., Influence du niveau d'humidité de l'air sur le sommeil du nouveau-né en incubateur, RBM-News, 21, 9, 1999. Crossref

  13. Tourneux P., Libert J.-P., Ghyselen L., Léké A., Delanaud S., Dégrugilliers L., Bach V., Échanges thermiques et thermorégulation chez le nouveau-né, Archives de Pédiatrie, 16, 7, 2009. Crossref

  14. Caputa Michał, Rogalska Justyna, Nowakowska Anna, Effect of temperature on postanoxic, potentially neurotoxic changes of plasma pH and free iron level in newborn rats, Brain Research Bulletin, 55, 2, 2001. Crossref

  15. Thomas Karen A., Infant Weight and Gestational Age Effects on Thermoneutrality in the Home Environment, Journal of Obstetric, Gynecologic & Neonatal Nursing, 32, 6, 2003. Crossref

  16. Yoo Seoung-Chool, Kim Young-Ho, Kwon Chi-Ho, Flow visualizations and hot-wire measurements on air flow in two different neonate incubators, KSME International Journal, 15, 7, 2001. Crossref

  17. Delanaud Stéphane, Decima Pauline, Pelletier Amandine, Libert Jean-Pierre, Durand Estelle, Stephan-Blanchard Erwan, Bach Véronique, Tourneux Pierre, Thermal management in closed incubators: New software for assessing the impact of humidity on the optimal incubator air temperature, Medical Engineering & Physics, 46, 2017. Crossref

  18. Antonucci Roberto, Porcella Annalisa, Fanos Vassilios, The infant incubator in the neonatal intensive care unit: unresolved issues and future developments, Journal of Perinatal Medicine, 37, 6, 2009. Crossref

  19. Bach Véronique, Delanaud Stéphane, Barcat Lucie, Bodin Emilie, Tourneux Pierre, Libert Jean-Pierre, Distal skin vasodilation in sleep preparedness, and its impact on thermal status in preterm neonates, Sleep Medicine, 60, 2019. Crossref

  20. Oldenburg Neto C. F., de Amorim Mardson Freitas, Influence of water vapour pressure on the thermal comfort of premature newborn, in World Congress on Medical Physics and Biomedical Engineering 2006, 14, 2007. Crossref

  21. Bach Véronique, Abbiss Chris R., Libert Jean-Pierre, McCabe Susan M., Skin Temperatures of Back or Neck Are Better Than Abdomen for Indication of Average Proximal Skin Temperature During Sleep of School-Aged Children, Frontiers in Psychiatry, 11, 2020. Crossref

  22. Tourneux P., Libert J.-P., Le réchauffement du nouveau-né prématuré : historique des évolutions technologiques des incubateurs fermés, Perfectionnement en Pédiatrie, 3, 4, 2020. Crossref

  23. Bach Véronique, Libert Jean-Pierre, Thermoregulation and Metabolism, in Pediatric Sleep Medicine, 2021. Crossref

  24. De La Fuente L, Campbell D E, Rios A, Grieg A, Graff M, Brion L P, Frequency analysis of air and skin temperature in neonates in servo-controlled incubators, Journal of Perinatology, 26, 5, 2006. Crossref